Your location:Home > Products > Products
Products

YCYJS Insulating Oil Dielectric Loss And Resistivity Tester

Views:

 1. Overview
Automatic precision oil dielectric loss design and manufacture according to GB/ t5654-2007 "measurement of relative capacitance, dielectric loss factor and dc resistivity of liquid insulation materials".Used for measuring dielectric loss factor and dc resistivity of insulating liquid dielectric such as insulating oil.Integrated structure.Internal integration of dielectric loss oil cup, temperature controller, temperature sensor, dielectric loss test bridge, ac test power supply, standard capacitor, high resistance meter, dc high voltage source and other main components.The instrument is equipped with a 5.7-inch TFT pure color LCD touch screen and a Chinese menu. The test results can be stored and printed out automatically. The operator can use it skillfully without professional training.

2.The main functions and characteristics
Oil cup of the accord with national standard GB/T5654-2007 three electrode type structure, the electrode spacing 2 mm,Can eliminate the stray capacitance and xie's influence on the dielectric loss test results.
Instrument by medium frequency induction heating, PID temperature control algorithm.The heating mode with oil cup with non-contact heating body, heat evenly, high speed, convenient control,Make temperature strictly controlled within the preset temperature error range.
Internal standard capacitor for SF6 inflatable three pole type capacitor, the capacitance of the dielectric loss and capacitance is not affected by environmental temperature, humidity,The instrument accuracy after long time use is still guaranteed.
Test AC power supply adopts AC - DC - AC conversion method,Effectively avoid the mains voltage and frequency fluctuations impact on dielectric loss test accuracy,Even if is a generator, the instrument can also run correctly.
Perfect protection function.When over voltage, over current, high voltage short circuit,Instrument can quickly cut off the high pressure, and send out a warning message.. When the temperature sensor failure or not connected, the warning information.In the medium frequency induction heating furnace with finite temperature relay,When the temperature over 120 degrees, the relay release, stop heating.
Parameters setting is convenient.Temperature set range of 40 ~ 120 ℃,Ac voltage setting range200 ~ 2200 v,0 ~ 500 v dc voltage setting range.
Adopted the pure color TFT LCD touch screen display, show clear.Simply set, the instrument can automatically test.And automatic storage and print the test results.
Built-in real-time clock, the test date, time may be retained with the test results, display and print.
Empty cup electrode calibration function.Measure the empty cup electrode capacitance and dielectric loss factor, cleaning and assembly of electrode to determine empty cup.Calibration data is automatically saved for relative permittivity and precise calculation of dc resistivity.
Instrument RAM9 platform as the core, high precision, high speed.

3.The main technical indicators

The power supply voltage: AC 220V±10%
Power frequency: 50Hz/60Hz ±1%
Measuring range:Capacitance5pF~200pF
The relative permittivity:1.000~30.000
Dielectric loss factor:0.00001~100
The dc resistivity:2.5 MΩm~20 TΩm
Accuracy of measurement: Capacitance±(1%reading+0.5pF)
The relative permittivity±1%reading
Dielectric loss factor±(1%reading+0.0001)
The dc resistivity±10%reading
Points and distinguishing rate:capacitance 0.01pF
The relative permittivity0.001
Dielectric loss factor0.00001
Temperature measurement range:40~120℃
Temperature measurement error: ±0.5℃
The ac voltage:200~2200VContinuous adjustable, frequency 50 hz
The dc test voltage: 0~500VContinuous adjustable
Power consumption:100W
Type or outside: 460mm× 370mm×330mm
Total amount of: 25Kg

4.Conditions of use
The environment temperature: 0℃~40℃
Relative humidity: <75%

Back To Top

Mobile station QR code
Copyright © Baoding Yuanchuang Electric Power Co., Ltd. All Rights Reserved